EXTERNAL PACEMAKERS

Pace 300®

Triple Chamber External Pacemaker

PACE 300
Oscor’s PACE 300 is the world’s first temporary three-chamber cardiac pacemaker featuring bi-ventricular pacing capability with individually programmable ventricular pacing and sensing channels and programmable inter-ventricular stimulation delay. It is designed for cardiac resynchronization therapy with one atrial and two independent ventricular channels. The available trigger modes and the adjustable interventricular stimulation delay provide a resynchronization option of both ventricles for different heart diseases: Arrhythmia and cardiac conduction disorders, Bradycardia after cardiac surgery, Pacemaker implant or replacement procedures.

Features

  • Easy setup of stimulation parameters by ergonomically designed dials and soft keys
  • High rate function for rapid atrial pacing of up to 1,000 ppm (adjustable while applying rapid atrial pacing therapy)
  • Activate emergency stimulation mode (D00, 12 V, 80 ppm, BV) by pressing “EMERGENCY” button
  • Convenient measurement of P/R wave peak amplitudes while pressing “PAUSE” button
  • PVARP and MTR depending on selected stimulation rate
  • Optional AUTO SENSE function which automatically tracks P/R wave peak amplitudes and adjusts atrial and ventricular sensitivities accordingly
  • Automatic mode switch to asynchronous modes upon detection of interference
  • Continuous surveillance of battery exhaustion and pacing lead system integrity
  • Large backlit display informs quickly about battery status, pacing mode, parameter settings, activated automatic functions and lead system safety surveillance
  • LEDs indicate whether knobs for ventricular output and sensitivity apply to RV + LV, RV only or LV only
  • Customize, save and recall pre-set parameter values for quick setup
  • Multilingual menu options, including English, Spanish, German, French, Italian or Swedish
  • Standard 9 V Alkaline battery provides at least 8 days of continuous operation (DDD, standard parameters)
  • Optional AC power supply saves battery power. Battery must remain in battery compartment in case of AC power shortage
  • Patient cable system featuring patented connections with protected pins for convenient, reliable electrical connections to increase patient safety
  • Standard 2 year limited warranty

TECHNICAL SPECIFICATIONS+

Set includes:

    Standard 9V battery, 3 ATAR extension cables, and a carrying case.

Stimulation Rate 30 ... 220 ppm
Maximum Tracking Rate 80 ... 230 ppm
Atrial overdrive stimulation 70 ... 1000 ppm
Stimulation amplitude 0.1 ... 18 V ± 10%
Pulse duration 0.05 ... 1.5 ms ± 5%
Sensitivity Atrium: 0.2 ... 20 mV ± 10%, infinite, Ventricle: 1.0 ... 20 mV ± 10%, infinite
Refractory Period 250 ... 400 ms 100 ... 500 ms 250 ms
A-V delay 5 ms ... 400 ms ± 5% ± 2 ms
V-V Stimulation Delay VVD 5 ms ... 80 ms ± 5% ± 2 ms
Pacing modes DDD (BV), DDD (LV), DDD (RV), VVI (BV), VVV (LV), VVI (RV), AAI, VDD (BV), VDD (RV), VDD (LV), DDT, DDI, DVI, DAI, DDD +AT, DAT, DOO, ODO, VVT, VOO, AAT, AOO, VDI, VAT
PACE 300